New NMR Commands
On OUAL

To start up login in to oual
 

Login: carter

Password: (see Don Carter)

Type:startx

click on NMR on the TASKBAR



Command                      Notes:

up # steps                      increase the frequency

                                      1 step is ~0.5 gauss

                                      Full range is ~5000 steps

down # steps                  reduce the frequency

nmrwhere type               current frequency
 

nmrgoto <frequency>     move NMR to frequency in integer or real Mhz

findnmr                            read the magnet current and move the nmr to the calculated frequency
 

browse #steps [delay]     cycle NMR from

                                            current value - [# steps/2]

                                          current value + [# steps/2]

                                          optional delay

                                          400 = default ~8 steps/sec

                                           800 = ~4 steps/sec

                                          Do not use less than 50 for delay
 

lock                                 Show NMR position continuously

Analysing Magnet Current (AMP) = 3.93682*(NMR in MHz)